Abstract
This paper reports the development and characterisation of an assembly technology for a polymer lab-on-chip. The system consists of a 150 μm deep hot embossed microfluidic channel in polycarbonate and Au electrodes fabricated separately by photolithography on polyethylenenaphthalate. The system is designed for impedimetric immunoassay detection in whole blood. Electrode layer and microfluidic substrate are joined by means of a 50 μm thick double-sided medical grade adhesive tape, adjusted with an optical alignment system. The bond proved to be liquid tight at room temperature. An alignment accuracy of 34 μm (+/- 19 μm) evaluated over a set of 23 samples, was achieved. The effect of alignment accuracy of the intermediate adhesive film on whole blood flow properties in the device is studied. Already an alignment error of 70 μm increases the flushing out time of whole blood by approximately 20%.
Original language | English |
---|---|
Pages | 1004-1009 |
DOIs | |
Publication status | Published - Jun 2010 |
Event | IEEE 60th Electronic Components and Technology Conference - Las Vegas, United States Duration: 1 Jun 2010 → 4 Jun 2010 |
Conference
Conference | IEEE 60th Electronic Components and Technology Conference |
---|---|
Abbreviated title | ECTC 2010 |
Country/Territory | United States |
Period | 1/06/10 → 4/06/10 |